Sound and Vibration Instrument Hire and Calibration
B&K 2250 BZ 7223 Frequency Analysis Specifications
To see the specifications for the various options available, click a link

BZ 7222 : Sound Analysis Software
BZ 7223 : Frequency Analysis Software
BZ 7224 : Logging Software
BZ 7225 : Enhanced Logging Option
BZ 7226 : Sound Recording Software
BZ 7227 : Reverberation Time Option
BZ 7228 : Sound Insulation Option
BZ 7230 : FFT Analysis Option
BZ 7231 : Tone Assessment Option
B&K 2250 : Home Page
B&K 2250 : Specifications
B&K 2250 : Detailed Image
BZ 7223 : Frequency Analysis Software
BZ 7224 : Logging Software
BZ 7225 : Enhanced Logging Option
BZ 7226 : Sound Recording Software
BZ 7227 : Reverberation Time Option
BZ 7228 : Sound Insulation Option
BZ 7230 : FFT Analysis Option
BZ 7231 : Tone Assessment Option
B&K 2250 : Specifications
B&K 2250 : Detailed Image
Bruel & Kjaer 2250 : Frequency Analysis Software - BZ 7223 | included |
---|
Conforms with BS EN 61260-1:2014
Centre Frequencies
1/1-octave Band Centre Frequencies: 8Hz to 16 kHz
1/3-octave Band Centre Frequencies: 6.3Hz to 20 kHz
Measurements
X = frequency weightings A, B, C or Z
Y = time weightings F or S
Data for Storage
Full Spectral Statistics
Spectra for Display and Storage
LXeq | LXSmax | LXFmax |
LXSmin | LXFmin |
Spectra for Display Only
LXS | LXF | LXYN1 |
LXYN2 | LXYN3 | LXYN4 |
LXYN5 | LXYN6 | LXYN7 |
SIL | PSIL | SIL3 |
---|---|---|
LXeq(f1-f2)a | ||
NR | NR Decisive Band | |
RC | RC Classification | |
NCB | NCB Classification | |
Loudness | Loudness Level |
Measuring Ranges
When using a Microphone Type 4189
Dynamic Range: From typical noise floor to max. level for a pure tone signal at 1 kHz 1/3-octave: 1.7 to 140 dB
Linear Operating Range: In accordance with IEC 61260 <20.5 dB to 140 dB.
Sampling for Octave or 1/3-octave Statistics
X = frequency weightings A or B
The Statistics can be based on either LXF or LXS:
Statistics LXFN1-7 or LXSN1-7 are based on sampling LXF or LXS, respectively, every T ms into 1 dB wide classes over 150 dB;
T = 100 for frequency range set to 12.5 - 20 kHz
T = 200 for frequency range set to 6.3 - 20 kHz
Full distribution can be saved with measurement
Measurement Displays
Spectrum: One or two spectra superimposed + A/B and C/Z broadband bars
Y-axis: Range: 5, 10, 20, 40, 60, 80, 100, 120, 140 or 160 dB. Auto zoom or auto scale available
Cursor: Readout of selected band
Cursor:
Readout of selected bandInternal Generator
Built-in pseudo-random noise generator
Spectrum: Selectable between Pink and White
Crest Factor:
Pink Noise: 4.4 (13 dB)
White Noise: 3.6 (11 dB)
Bandwidth: Selectable
Lower Limit: 50 Hz (1/3-oct.) or 63 Hz (oct.)
Upper Limit: 10 kHz (1/3-oct.) or 8 kHz (oct.)
Output Level: Independent of bandwidth
Max: 1 Vrms (0 dB)
Gain Adjustment: -60 to 0 dB
When bandwidth is changed, the level for all bands is automatically adjusted to comply with the set output level
Correction Filters: for sound sources Type 4294, Type 4295 and Type 4296: Flat or Optimum.
Repetition Period: 175 s
Output Connector: Output Socket
External Generator
Selectable as alternative to Internal generator
To control external noise generator, set:
Levels: 0 V (Generator off), 4.5 V (Generator on)
Rise-time and Fall-time: 10 µs
Escape Time: 0 to 60 s
Build-up Time: 1 to 10 s
* Our standard hire charges are based on a 4-day minimum period, details other options.