Gracey & Associates   01234 708835



Gracey Logo

Sound and Vibration Instrument Hire and Calibration

Bruel & Kjaer 2240 Sound Survey Meter Specifications


 
Bruel & Kjaer 2240 Noise Survey Meter
B&K 2240 : Sound Level - Precision Class 1 £ 10 a day *
 

Standards :

- conforms to the following:
IEC 61672-1:2002 Class 1
IEC 60651 Type 1 (1979) with amendments 1 and 2
IEC 60804 Type 1 (2000)
ANSIS1.4 - 1983 Type S1
ANSIS1.43 - 1997 Type 1
 

Parameters

- Four parameters are measured simultaneously:
LAF, instantaneous sound pressure level, frequency weighting 'A' and time weighting 'F'
LAFmax, maximum sound pressure level, frequency weighting 'A' and time weighting 'F'
LAeq, equivalent continuous sound pressure level, frequency weighting 'A'
LCpeak, maximum peak sound pressure level, frequency weighting 'C'
Resolution: 0.1dB for all four parameters
 

Measurement Control

Measurements are manually controlled
Measurement times between 1 s and 60 min
 

Measuring Ranges


RMS: Total range: 30 - 140 dB
Two manually selected ranges: 30 - 110 dB and 60 - 140 dB
Peak: 60 - 143 dB
 

Noise Floor

Below measurement range < 22 dB
 

Detectors


Simultaneous RMS and Peak with independent frequency weightings
Linear Operating Range: 80 dB
 

Frequency Range


20 Hz to 16 kHz
 

Frequency Weightings


A-weighting (RMS)
C-weighting (Peak)
 

Time Weightings


'F' Fast
 

Overload and Under-Range Indicators


Overload is indicated at full-scale +0.3 dB
An overload detected during a timed measurement latches and is displayed with the relevant parameters until these are cleared
Under-range is indicated at scale lower limit -0.5 dB
 

Microphone


Type 4188 Pre-polarized Free-field 1/2" Condenser Microphone Sensitivity: -30 dB re 1 V/Pa ± 2 dB (corresponding to 31.6 mV/Pa)
Frequency Range: 8 Hz to 16 kHz ±2 dB
 

Display


LCD showing:
Input signal level – indicated with a quasi-analogue bar
Selected parameters with level
Warnings for: low battery, measurement in progress, overload and under-range
Measuring range
Frequency and time weighting
Elapsed measurement time
Update Cycle: 1 s (numeric), 0.1 s (bar graph)
Elapsed Time Display: from 00 min 00 s up to 60 min 00 s
 

Settling Time


Less than 5 s
 

Reference Conditions


Reference Frequency: 1000 Hz
Reference Sound Pressure Level: 94 dB
Reference Temperature: 20 °C (68 °F)
Reference RH: 65%
Reference Range: 30-110 dB
Reference Direction of Incidence: Frontal
 

Environmental Effects


Storage Temperature: -25 to 60 °C (-13 to 140 °F)
Operating Temperature: -10 to 50 °C (14 to 122 °F)
Effect of Magnetic Field: 80 A/m (1 oersted) at 50 Hz gives less than 30 dB (A-weighted)
 

Batteries


Two 1.5 V Lr6/AA-size alkaline batteries
Power Consumption During Normal Operation: Less than 300 mW
Lifetime (at room temperature): Approximately 16 hours
 

Physical Characteristics


Dimensions: 230 x 78 x 31 mm including microphone
 

Weight


245 g (8.64 oz) including batteries
 
E&OE - specifications subject to change without notice.
* Our standard hire charges are based on a 4-day minimum period, details other options.